Test Utilization

Influenza Testing Update: April 2024

This winter, West Michigan influenza infections peaked in February 2024 and significantly declined throughout March and April 2024. Now with a low prevalence of circulating influenza, using the most sensitive diagnostic test is recommended in order to obtain accurate and actionable results. Influenza antigen testing (e.g. Sofia instrument) is no longer appropriate at this time and Influenza PCR testing should be ordered when clinically indicated.

 

Low Influenza
Prevalence
High Influenza
Prevalence
Recommended Order Influenza PCR (LAB3255) Influenza antigen testing (e.g. Sofia instrument)
or
Influenza PCR (LAB3255)

More detailed information can be found as published by the CDC:

Algorithm to assist in the interpretation of influenza testing results and clinical decision-making during periods when influenza viruses are NOT circulating in the community, and in previous lab newsletters: Influenza Diagnosis and Test Utilization

Test Utilization

Influenza Testing Recommendation (December 2023)

Influenza cases numbers are currently increasing throughout West Michigan. Influenza prevalence guides the most appropriate testing approach and sufficient prevalence has now been reached for influenza rapid antigen or point of care testing (e.g. Sofia instrument) to have improved performance. The use of point of care testing as a screening method during high prevalence reduces both the cost to the patient and turn-around-time as compared to lab-based molecular methods. Point of care tests generally have high analytical specificity, but lack the sensitivity of nucleic acid amplification methods (e.g. PCR). For this reason, PCR testing is still recommended for hospitalized patients and for outpatients with a negative result if influenza is still suspected and if the result will impact clinical decision making.

Low influenza prevalence High influenza prevalence
Recommended order Influenza PCR  Influenza Point of Care Testing
or
Influenza PCR 

Continue Reading

Test Utilization

Influenza Testing Update: April 2023

This winter, West Michigan influenza infections peaked in December 2022 and significantly declined throughout February and March 2023. Now with a low prevalence of circulating influenza, using the most sensitive diagnostic test is recommended in order to obtain accurate and actionable results. Influenza antigen testing (e.g. Sofia instrument) is no longer appropriate at this time and Influenza PCR testing should be ordered when clinically indicated. Continue Reading

General Information, Test Utilization

Influenza Testing Recommendation (November 2022)

Influenza cases numbers are currently increasing throughout West Michigan. Influenza prevalence guides the most appropriate testing approach and sufficient prevalence has now been reached for influenza rapid antigen testing (e.g. Sofia instrument) to have improved performance. The use of rapid antigen testing as a screening method during high prevalence reduces both the cost to the patient and turn-around-time as compared to lab-based molecular methods. Rapid antigen tests generally have high analytical specificity, but lack the sensitivity of nucleic acid amplification methods (e.g. PCR). For this reason, PCR testing is still recommended for hospitalized patients and for outpatients with a negative antigen result if influenza is still suspected and if the result will impact clinical decision making. Continue Reading

General Information

Regionals New Test Communication: “4-Plex”

Effective immediately, the following Regional Hospital Laboratories have the 4-Plex (COVID-19, Influenza A, Influenza B, RSV) test available for ordering and testing:

  • Big Rapids Hospital Laboratory
  • Blodgett Hospital Laboratory
  • Gerber Hospital Laboratory
  • Ludington Hospital Laboratory
  • Pennock Hospital Laboratory
  • Reed City Hospital Laboratory
  • United/Greenville Hospital Laboratory
  • Zeeland Hospital Laboratory

Continue Reading

General Information, Test Spotlight

Influenza Testing Update: April 2022

West Michigan influenza infections peaked mid-March 2022 and have been declining in the weeks that followed. Now with a low prevalence of circulating influenza, using the most sensitive diagnostic test is recommended in order to obtain accurate and actionable results. Influenza antigen testing (e.g. Sofia instrument) is no longer appropriate at this time and Influenza PCR testing should be ordered when clinically indicated.

Low Influenza
Prevalence
High Influenza
Prevalence
Recommended Order Influenza PCR (LAB3255) Influenza Rapid Antigen (LAB2111530)
or
Influenza PCR (LAB3255)

 

More detailed information can be found as published by the CDC:

Algorithm to assist in the interpretation of influenza testing results and clinical decision-making during periods when influenza viruses are NOT circulating in the community, and in previous lab newsletters: Influenza Diagnosis and Test Utilization

 

TEST INFORMATION

Test name Epic code Interface EMR Code CPT Code
 Influenza PCR  LAB3255 11594 87502
 Influenza Rapid Antigen  LAB2111530 11208 87804 x4
Compliance & Safety, Education & Training, General Information

Appropriate Labeling for Shared Specimen – COVID-19 & FLU AB PCR

To provide efficient and safe testing for patient specimens, please review the following information when sending COVID-19 PCR and Influenza A/B PCR testing in the same encounter to Spectrum Health Laboratories.

Two labels will be available for this testing. Labels must be fixed so that

• The test and patient information are shown
• The barcode may be easily scanned by Lab staff or instruments

When this is not done correctly, Laboratory staff removes the labels and re-fixes or re-prints it, which obstructs workflow, causes delays, and could result in patient information being peeled off. With patient safety and efficient testing being two of the Lab’s top priorities, it is vital that specimens arrive in such a way as to best accomplish both. Continue Reading

Microbiology, Test Spotlight

Influenza Testing Recommendation Update (December 2021)

The mitigation strategies enacted during the COVID-19 pandemic to reduce the spread of this virus led to essentially non-existent influenza prevalence during the 2020-2021 winter season. However, influenza has now returned and case numbers are increasing throughout West Michigan.

Influenza prevalence guides the most appropriate testing approach and sufficient prevalence has now been reached for influenza rapid antigen testing (e.g. Sofia instrument) to have improved performance. The use of rapid antigen testing as a screening method during high prevalence reduces both the cost to the patient and turn-around-time as compared to lab-based molecular methods. Rapid antigen tests generally have high analytical specificity, but lack the sensitivity of nucleic acid amplification methods (e.g. PCR). For this reason, PCR testing is still recommended for hospitalized patients and for outpatients with a negative antigen result if influenza is still suspected and if the result will impact clinical decision making. Continue Reading

Advanced Technology Laboratory, General Information, Main lab, Test Utilization

Outpatient Respiratory Viral Testing

Spectrum Health Lab has noticed an increase in outpatient orders for Respiratory Pathogens by Film Array [LAB3359].  This may be due to a shortage of supplies for some Point of Care (POC) testing platforms.  Film Array is typically used for emergency and high acuity patients and therefore has a high cost that may not be covered by most patients’ insurance.  To lesson out of pocket costs for your patients, please order the below for RSV, COVID, or Flu testing. Continue Reading

Main lab, Microbiology, Test Utilization

Influenza Testing Reminder

The mitigation strategies enacted during the COVID-19 pandemic to reduce the spread of this virus have also impacted the transmission of other respiratory viruses. Influenza is typically prevalent during the winter months (December through March), however, influenza cases were essentially non-existent during the 2020-2021 season and influenza is not currently circulating in our community.

In the absence of circulating influenza activity, it is recommended to use the most sensitive diagnostic test in order to obtain accurate and actionable results. Influenza antigen testing (e.g. Sofia instrument) is not appropriate at this time. More detailed information can be found as published by the CDC: Algorithm to assist in the interpretation of influenza testing results and clinical decision-making during periods when influenza viruses are NOT circulating in the community

Low Influenza
prevalence
High influenza
prevalence
Recommended order Influenza PCR (LAB3255) Influenza Rapid Antigen (LAB2111530)
or
Influenza PCR (LAB3255)

TEST INFORMATION

Test name Epic code Interface EMR Code CPT Code
 Influenza PCR  LAB3255 11594 87502